FORS2 Grism Efficiencies

GRIS_1400V+18

This grism is based on a Volume Phased Holographic Grating

4000.10
4100.16
4250.27
4500.49
4750.71
5000.83
5250.87
5500.80
5750.67
6000.49
6250.32
6500.18

GRIS_1200R

This grism is based on a Volume Phased Holographic Grating

540     13.9
550 20.46
560 28.38
570 36.85
580 45.61
590 54.5
600 63.11
610 70.66
620 77.62
630 83.29
640 87.67
650 90.3
660 91.4
670 91.28
680 89.35
690 85.81
700 81.01
710 75.23
720 68.68
730 61.84
740 53.88
750 46.45
760 38.8
770 32.03
780 24.99
790 18.93
800 12.96

GRIS_1028z+29

This grism is based on a Volume Phased Holographic Grating

nm efficiency

7500.43
7750.59
8000.74
8250.84
8500.89
8740.89
9000.83
9250.72
9500.59
9750.44
9800.41

GRIS_600B+22

nm efficiency

330 0.55
350 0.63
375 0.67
400 0.68
450 0.61
500 0.53
550 0.46
600 0.37
650 0.32
700 0.26

GRIS_600R+24

nm efficiency

425 0.38
450 0.43
475 0.48
500 0.50
550 0.52
600 0.49
650 0.45
700 0.42
750 0.37
800 0.34
850 0.31
900 0.27
950 0.23

GRIS_600RI+19

This grism is based on a Volume Phased Holographic Grating

nm efficiency

4850.29
5000.32
5250.38
5500.45
5750.51
6000.58
6250.65
6500.71
6750.77
7000.82
7250.86
7500.87
7750.86
8000.85
8250.81
8500.76
8750.70
9000.62
9250.54
9500.45
9750.36
10000.28

GRIS_600I+25

nmefficiency

450 0.28
500 0.43
550 0.54
600 0.57
650 0.56
700 0.55
800 0.47
900 0.41
1000 0.34

GRIS_600z+23

This grism is based on a Volume Phased Holographic Grating

nmefficiency

8000.69
8250.76
8500.82
8750.86
9000.88
9250.88
9500.86
9750.84
10000.80

GRIS_300V+20

nmefficiency

330 0.48
350 0.62
375 0.72
400 0.78
450 0.79
500 0.74
550 0.63
600 0.55
650 0.48
700 0.42
750 0.37
800 0.33
850 0.29
900 0.25

GRIS_300I+21

nm efficiency

500 0.14
550 0.31
600 0.47
650 0.58
700 0.66
750 0.69
800 0.70
850 0.67
900 0.65
950 0.61
1000 0.57
1050 0.53
1100 0.50
1150 0.43

GRIS_200I+28

nm efficiency

550 0.10
560 0.33
580 0.56
600 0.67
650 0.75
700 0.78
750 0.77
800 0.75
900 0.67
1000 0.58
1100 0.50

GRIS_150I+27

nm efficiency

340 0.30
360 0.44
375 0.53
400 0.66
450 0.78
500 0.80
550 0.76
600 0.70
650 0.62
700 0.56
800 0.44
900 0.35
1000 0.28
1100 0.22

EGRIS_110B+98

Echelle grism

4th order
---------
nm efficiency

750 0.05
800 0.14
825 0.20
850 0.24
875 0.27
900 0.29
925 0.29
950 0.27
975 0.25
1000 0.21


5th order
---------
nm efficiency

625 0.05
650 0.14
675 0.21
700 0.28
725 0.29
750 0.30
775 0.28
800 0.26
825 0.23
850 0.18
875 0.15
900 0.11
925 0.08


6th order
---------
nm efficiency

550 0.07
575 0.16
600 0.25
625 0.27
650 0.25
675 0.22
700 0.17
725 0.14
750 0.09
775 0.07
800 0.05


7th order
---------
nm efficiency

512 0.15
525 0.18
537 0.21
550 0.24
562 0.26
575 0.23
600 0.18
625 0.13
650 0.08


8th order
---------
nm efficiency

450 0.09
462 0.14
475 0.21
481 0.23
488 0.26
494 0.26
500 0.26
506 0.25
512 0.23
525 0.20
550 0.11
575 0.05


9th order
---------
nm efficiency

400 0.02
412 0.08
425 0.19
438 0.25
450 0.26
462 0.23
475 0.18
500 0.07
525 0.02


10th order
----------
nm efficiency

375 0.05
388 0.17
400 0.25
412 0.25
425 0.18
438 0.13
450 0.09
475 0.02


11th order
----------
nm efficiency

350 0.08
362 0.18
375 0.25
388 0.20
400 0.16
412 0.08
425 0.03


12th order
----------
nm efficiency

332 0.06
338 0.11
344 0.18
350 0.21
356 0.22
362 0.22
375 0.14
388 0.06

XGRIS_600B+92

Cross disperser grism for the echelle EGRIS_110B+98

nm efficiency

325 0.25
350 0.47
375 0.58
400 0.65
450 0.66
500 0.63
550 0.56
600 0.48
700 0.35
800 0.25