FORS Filter page

The FORS instruments can accommodate
  • Broadband filters in the collimated beam: There are 3 wheels in the collimated beams with 7 positions available per wheel. A number of positions are usually occupied by grisms and (in FORS1) the Wollaston prism. Broadband filters have a free diameter of 138mm and are bayonet mounted.
  • Interference filters in the converging beam of the camera: There are 2 wheels in the converging beam with together 8 available positions. Interference filters have a size of 115mm and are bayonet mounted.

FORS1 Standard Filters

Presently the broadband filters listed in the following table are available for FORS1. In addition, the interference filters listed below can be requested.

Filter name+ESO number eff. wavelength (nm) FWHM (nm)
u_HIGH+112 361 50.5 (1)
b_HIGH+113 440 103.5 (2)
v_HIGH+114 557 123.5
g_HIGH+115 470 160.5
U_BESS+33 (3) 336 36.0
B_BESS+34 (3) 429 88.0
V_BESS+35 554 111.5
R_BESS+36 657 150.0
I_BESS+37 768 138.0
u_GUNN+38 (3) 359 33.5
v_GUNN+39 (3) 398 46.0
g_GUNN+40 (4) 506 79.5
r_GUNN+41 (3) 653 81.5
z_GUNN+42 (3) 910 130.5
GG375+30 (5) n/a (edge filter) n/a
GG435+31 (5) n/a (edge filter) n/a
OG590+72 (5) n/a (edge filter) n/a

(1) red leak <5*10^-6
(2) red leak <1*10^-5
(3) Available on request. These filters do not form part of the standard instrument configuration. They can be inserted in the instrument as needed in the same manner as for interference filters.
(4) this filter is located in one of the interference filter wheels, as it is designed as an interference filter
(5) these are intended as order separation filters for spectroscopy

FORS2 Standard Filters

Presently the broadband filters listed in the following table are available for FORS2. In addition, the interference filters listed below can be requested.

Filter name+ESO number eff. wavelength (nm) FWHM (nm)
U_SPECIAL+73 362 29.0 (1)
B_BESS+74 429 88.0 (2)
V_BESS+75 554 111.5
R_SPECIAL+76 655 165.0
I_BESS+77 768 138.0
z_GUNN+78 910 130.5
GG375+80 (3) n/a (edge filter) n/a
GG435+81 (3) n/a (edge filter) n/a
OG590+32 (3) n/a (edge filter) n/a
FILT_465_250+82 (3) 465 250

(1) red leak <10^-5
(2) red leak <4*10^-4
(3) these are intended as order separation filters for spectroscopy

Interference Filters for FORS1 and FORS2

The following table lists all presently available interference filters. They are used for FORS1 and for FORS2. Their characteristics are given with the FORS SR and HR collimators: central wavelength, peak transmission and FWHM, all in nm. Due to their location in the converging beam, the filter characteristics depend on the collimator used. The filter bandwidths are wider, the central wavelength is blue-shifted, and the peak transmission is lower than in a parallel beam. With the SR collimator, the effect is larger than with the HR collimator. The filters are centered on important emission lines and on 5 and 10 % longer wavelengths.

Line filter name and ESO nr centr. wavelength T0 W0
Collimator   SR HR SR HR SR HR
  OII+44 371.7 372.9 0.45 0.48 7.3 6.9
[OII] 372.7 OII/4000+45 377.6 378.8 0.37 0.40 6.5 6.1
  OII/8000+46 381.4 382.6 0.43 0.47 6.5 6.1
  HeII+47 468.4 469.1 0.79 0.82 6.6 6.4
HeII 468.6 HeII/3000+48 472.6 473.4 0.76 0.79 5.8 5.6
  HeII/6500+49 478.1 478.9 0.78 0.81 6.8 6.6
  OIII+50 500.1 500.9 0.76 0.80 5.7 5.5
[OIII] 500.7 OIII/3000+51 504.5 505.3 0.76 0.80 5.9 5.7
  OIII/6000+52 510.5 511.3 0.74 0.78 6.1 5.9
  HeI+53 586.6 587.6 0.79 0.84 6.0 5.7
HeI 587.6 HeI/2500+54 592.0 593.0 0.77 0.81 6.8 6.5
  HeI/5000+55 597.5 598.5 0.85 0.89 7.4 7.2
  OI+56 629.5 630.6 0.75 0.79 7.2 6.9
[OI] 630.0 OI/2500+57 635.4 636.4 0.75 0.81 5.9 5.5
  OI/4500+58 640.4 641.4 0.77 0.83 6.3 6.0
  H_Alpha+83 656.3 657.4 0.70 0.76 6.1 5.7
H alpha 656.3 H_Alpha/2500+60 660.4 661.5 0.77 0.83 6.4 6.1
(2)  H_Alpha/4500+61 666.5 667.6 0.72 0.77 6.5 6.1
  SII+62 672.8 673.9 0.77 0.82 6.6 6.3
[SII] 672.4 SII/2000+63 677.4 678.5 0.77 0.82 6.8 6.5
  SII/4500+64 683.2 684.3 0.72 0.78 6.4 6.0
  SIII+65 952.3 953.9 0.68 0.80 5.9 5.2
[SIII] 953.2 SIII/1500+66 957.2 958.8 0.72 0.84 6.3 5.6
  SIII/3000+67 962.1 963.7 0.70 0.83 5.9 5.2
"special 485" (1) FILT_485_37+68 485   0.89   37  
"special 691" (1) FILT_691_55+69 691   0.93   55  
"special 815" (1) FILT_815_13+70 815   0.90   13  
"special 834" (1) FILT_834_48+71 834   0.90   48  
"special z" z_SPECIAL+43 915   0.94   20  
"special z" FILT_917_6+88 917   0.85   6  
  FILT_500_5+85 500   0.81   5  
[OIII] 500.7 FILT_503_5+86 503   0.83   5  
  FILT_530_25+84 530   0.85   25  

(1) night sky suppression filter

Measured Measured filters transmission curves

The broad band filter curves were measured on the optical bench. The transmission curves for many of the FORS interference filters have been measured within the instruments. The results are available as ASCII tables or ps-plots via this link

Non-standard filter policy

Users may propose to use their own filters with FORS in their applications for observing time. The following conditions apply:
  1. User supplied filters may only be used in Visitor Mode.
  2. ESO has now purchased a limited amount of filter mounts which will be sent to the users on request. To assemble the filters into the mount is under the full responsibility of the users.
  3. Filters supplied by users shall fulfill the following requirements:
    • Broadband filters are used in the collimated beam near the pupil. Their diameter shall not be smaller than 138mm to avoid vignetting which would be equivalent to a reduction of the main mirror diameter.
    • Interference filters (115mm) are used in the converging beam of the camera. Their spectral resolution can not exceed 100 (SR mode) or 400 (HR mode). More narrow interference filters will be convolved to lower resolution and low peak transmission due to the dependance of the transmission from the incident beam angle.
  4. More technical specifications as required to purchase the filters will be send to the users on request.
  5. The filters have to be send to Paranal about 6 weeks in advance for the final optical tests within the instrument and the software configuration.

example for interference filter specification:

Filter type: (at least) tri-cavity interference filter
CWL: 486.1nm+-1nm
FWHM: 6.0nm+-1nm
Minimum transmission: 80%
Used in: normal incidence
Incident beam: half cone angle of incident beam: 8.5 degrees
Size: 115mm+-0.25mm
Max thickness: 7.0mm+-0.25mm
Minmum clear aperture: 110mm diameter
Operating temperature: 0...+15C (nominal: 12C)
Storage temperature: -10C...+30C
Optical quality: image quality, transmitted wave front TWF: lambda/4 per inch (at central wave length), parallelism <30arcsec,Scratch-Dig: 60-40, minimum ghost, TWF of lambda/2 per 25mm
A/R coatings: anti-reflection coat all exterior surfaces, reflectivity <1.5 % per surface
Out-of-band blocking: 1E-4 UV to 1200nm
Edge: black-painted, no metal ring
Documentation: spectrophotometric trace of transmission in %T and of attenuation regions in OD, if possible data files in ascii or similar.

central wave lengths shifts - contributed filters:

Please note that the interference filters are installed into the converging beam of FORS and the central wavelength will shift bluewards in respect to the measured filter curves for a normal incident. Accordingly filters from other instruments are typically shifted in wavelength if used with FORS. Filters have to be purchased in most cases for the FORSes such that the central wavelength is right for the FORS instruments.