E2V CCD Test Reports
CCD nickname | 
Where | 
Rated by E2V as | 
Tested | 
Instrument Target | 
Rated by ODT as | 
Defects measured by ODT (E2V) | 
Notes | 
||
Traps (cols) (1)  | 
Bias hot cols (2)  | 
Dark hot cols. (3)  | 
|||||||
ENGINEERING-SETUP GRADE DEVICES | 
|||||||||
Alan
 | 
La Silla, 3.6m 
 | 
Eng device 
 | 
Jan 2000 
 | 
CES | 
Science 
 | 
71 
 | 
0 
 | 
0 
 | 
Outstanding QE, but large amount of traps (left side), has been shipped to Chile for CES final upgrade 
 | 
Bruce
 | 
ODT lab/Safe 
 | 
Eng device 
 | 
March 2000 
 | 
Giraffe | 
Science but avoid mosaics, Giraffe 
 | 
0 
(8)  | 
3 
(4)  | 
3 
 | 
Despite a 20 um overhang this is an outstanding device with high QE. Glowing on top 50 lines out. 
 | 
Ringo
 | 
Copenhagen 
 | 
Eng device 
 | 
Feb 2000 
 | 
DFosc | 
Classic Eng device, not so good, not so bad | 1 
(7)  | 
10 
(23)  | 
11 
 | 
Sent to Danish people (DFOSC) as FINAL CCD, in March 2000 
 | 
Julio
 | 
ODT lab/test cryostat 
 | 
Setup grade 
 | 
- 
 | 
None | 
Not tested 
 | 
- 
 | 
- 
 | 
- 
 | 
Mounted in Berta Head, Sebastian is playing with for contamination tests. 
 | 
Iggy
 | 
La Silla/ODT 
 | 
Setup grade 
 | 
- 
 | 
Spare 
 | 
Tested data to be gathered 
 | 
- 
 | 
- 
 | 
- 
 | 
Used for preliminary tests for CES upgrade, in Chile, should be shipped back in Garching at some point. 
 | 
Eva
 | 
ODT lab/Safe 
 | 
Eng device 
 | 
Sep 2000 
 | 
None | 
Seems dead 
 | 
- 
(4)  | 
- 
(34)  | 
- 
 | 
Given on loan to Danish people (DFOSC) in September 1999, should be shipped back in Garching at some point. Danish people may have killed this device. Tested here looks dead.... 
 | 
Rachel
 | 
ODT lab/Safe 
 | 
Eng device 
 | 
Aug 2000 
 | 
Omegacam
 | 
Could be used for imagery (Science) 
 | 
1 
(1)  | 
8 
(18)  | 
4 
 | 
Good imaging device, should be exchanged by a non frame transfer Omegacam CCD 
 | 
Satriani
 | 
ODT lab/Safe 
 | 
- 
 | 
None 
 | 
? 
 | 
? 
 | 
? 
 | 
? 
 | 
Likely an enginering device 
 | 
|
Antonio
 | 
ODT lab/Safe 
 | 
Eng device 
 | 
Sep 2000 
 | 
Omegacam
 | 
- 
 | 
- 
(17)  | 
- 
(18)  | 
- 
 | 
Used for weak contamination tests, should be exchanged by a non frame transfer Omegacam CCD 
 | 
SCIENCE GRADE DEVICES | 
|||||||||
Carreras
 | 
ODT testbench 
 | 
Science 
 | 
Oct 2007 
 | 
Giraffe | 
Science 
 | 
(1) 
 | 
(0) 
 | 
(0) 
 | 
|
Kate
 | 
ODT lab/Safe 
 | 
Science 
 | 
Feb 2000 
 | 
Spare | 
Science 
 | 
0 
(3)  | 
2 
(4)  | 
3 
 | 
This device has not been thoroughly tested due to charge injection from the wiping drain. VDD must be reduced. 
 | 
Celine
 | 
ODT lab/Safe 
 | 
Science 
 | 
Jan 2000 
 | 
Spare | 
Science Spare 
 | 
2 
(3)  | 
0 
(1)  | 
0 
 | 
2 sets of trap, perfect dark frame 
 | 
Jasmin
 | 
ODT lab/Safe 
 | 
Science 
 | 
Jan 2000 
 | 
Harps | 
Science, HARPS possible candidate 
 | 
1 
(7)  | 
0 
(1)  | 
? 
 | 
Very nice device, no defects 
 | 
Linda
 | 
ODT lab/Safe 
 | 
Science 
 | 
Dec 1999 
 | 
Harps | 
Science, HARPS possible candidate 
 | 
2 
(1)  | 
0 
(2)  | 
0 
 | 
Nice, no dark related defects 
 | 
Paul
 | 
ODT lab/Safe 
 | 
Science 
 | 
Jan 2000 
 | 
None | 
Science 
 | 
1 large  
(3)  | 
1 
(5)  | 
1 
 | 
Has an ugly trap a X=1500, 5 cols 
 | 
Barbarella
 | 
ODT lab/Safe 
 | 
Science 
 | 
Dec 1999 
 | 
None | 
Cannot be used for spectroscopy 
 | 
0 
(1)  | 
? 
(1)  | 
? 
 | 
Has a remanent defect in the middle, long dark frame to remake.Exchanged with BRUCE. 
 | 
Annie
 | 
ODT lab/Safe 
 | 
Science 
 | 
Feb 2000 
 | 
None | 
Science 
 | 
0 
(2)  | 
1 
(2)  | 
2 
 | 
Not so nice for a science grade...! Beware ! VDD (Drain) must be 11V and VDR = 24 V (otherwise massive Glowing will occur)  | 
Elvis
 | 
ODT lab/Safe 
 | 
Science 
 | 
Feb 2000 
 | 
Omegacam tracker spare | 
Could be used for imagery (Science) 
 | 
2 
(5)  | 
5 
(5)  | 
6 
 | 
Nice Eng grade device, It looks like that E2V has used a trick to reduce IR fringing, because IR fringing is different than usual. 
 | 
DEVICES SENT BACK TO E2V | 
|||||||||
Satriani v1
 | 
Sent back to E2V 
 | 
Eng device 
 | 
March 2000 
 | 
- 
 | 
Science, very nice 
 | 
1 
 | 
1 
 | 
1 
 | 
E2V said that this device has 82 traps ... it turns out to be a very nice science grade usable for spectroscopy... mistake from them ? Yes ! Indeed, this is 8462-13-2 instead of 8462-14-2, shipped back to E2V. 
 | 
Satriani v2
 | 
Sent back to E2V 
 | 
Eng device 
 | 
Aug 2000 
 | 
- 
 | 
Unusable 
 | 
(82) 
 | 
(103) 
 | 
- 
 | 
Device has a huge charge injection in serial register, always saturated, not usable at all, rejected ! 
Sent back to E2V around beginning of Sep 2000  | 
CCD nickname | 
Where | 
Rated by E2V as | 
Tested | 
Intrument Target | 
Rated by ODT as | 
Defects measured by ODT (E2V) | 
Notes | 
||
Traps (cols) (1)  | 
Bias hot cols (2)  | 
Dark hot cols. (3)  | 
|||||||
(1) trap visible with a 650 nm flat field
 (2) hots columns visible over a bias frame
 (3) hots columns visible over an hour dark exposure (including one visible in a bias frame)