XSHOOTER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2026-04-18T22:36:01 (UT)
  | now: 2026-04-18T23:23:33 (UT) muc02 QC pipeline: xshoo-3.6.5a (installed 2023-09-18)
same group: bias_AGC dark_AGC detmon_AGC
General news: NOTE: The Health Check and calChecker services have been moved to qcFlow. These pages are only maintained for comparison and will be removed in 2026.
XSHOOTER news:  
Report news:
DATE*: [?] 2026-04-11
report | NLT
2026-04-12
report | NLT
2026-04-13
report | NLT
2026-04-14
report | NLT
2026-04-15
report | NLT
2026-04-16
report | NLT
2026-04-17
report | NLT
2026-04-18
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.78e+03 ADU 20%± 356 95 median_master this | last_yr | all  
2 QC1DB MEDIAN 0.694 ADU 80%± 0.555 95 struct_x1 this | last_yr | all  
3 QC1DB MEDIAN 4.54 ADU 60%± 2.72 95 ron this | last_yr | all  
4 QC1DB MEDIAN 0.863 ADU 60%± 0.518 95 struct_y1 this | last_yr | all  
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:xshooter_bias
(QC1 database)
dataset:median_master
median:1.78e+03ADU
rel. thresholds: ±20%1424.3933592...2136.5900388ADU
N_data plotted:95
[click plot for closeup]
Plot 3
scores: 8|
data source:xshooter_bias
(QC1 database)
dataset:ron
median:4.54ADU
rel. thresholds: ±60%1.814882...7.259528ADU
N_data plotted:95
[click plot for closeup]
Plot 2
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_x1
median:0.694ADU
rel. thresholds: ±80%0.138795...1.249155ADU
N_data plotted:95
[click plot for closeup]
Plot 4
scores: not implemented
data source:xshooter_bias
(QC1 database)
dataset:struct_y1
median:0.863ADU
rel. thresholds: ±60%0.3450036...1.3800144ADU
N_data plotted:95
[click plot for closeup]
This plot

This plots shows detector related QC parameters derived from BIAS health check calibrations of the xshooter acquisition and guiding camera (AGC). It is a 562 x 528 pixel technical CCD. For the definition of the QC parameters, please see the corresponding section of the UVB-arm and VIS-arm science detector. On 2014-01-09 the detector operations mode was switched from DET.OUT1.GAININD = 8 to DET.OUT1.GAININD = 4.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )