2022-10-02 The kappa matrix has a validity of 30 days, moreover kappa matrix in MED tilted by 10-20 px does not affect the night observations. The IOT is monitoring the kappa so please ignore for the time being the flag, unless you see for several days in a raw more than 100 px shift. CPa [matisse@eso.org]
2022-12-13 templates are interrupted after only 1/2 the frames, red scores are not significant in this cases [qc_matisse@eso.org]
2022-02-20 Bug in the software. MATISSE Team is aware of this. Working on it.
2023-02-26: the issue seems solved : this was not an pipeline issue but a SHIFT MAP not updated after the switch of lamps (IPE) [ccid@eso.org]
2021-09-28 the maximum OPD is calculated on the 6 baselines. Specifications mentioned a maximum shift of 100 microns/8h for MATISSE-LM and 300 for MATISSE-N. F [qc_matisse@eso.org]